PROPRIETARY DATA: This document contains proprietary data of ABB Inc. No disclosure, distribution (electronic or otherwise),
or other means of dissemination may be made without written permission.
3BUS 208 000 R1301 139Measurement Setup
Table 6-22 Range for Basis Weight Sensor Standardize Limits
Note: You have to observe source decay, which will reduce the AIS value
by 3.4% per month for STLP and 0.7% per month for STLK sensors.
Host Report
Name
Variable Name Function
Acceptable
Values
AIZ AIZ Nominal
AIZ Range
Expected voltage with shutter
closed
Plus and minus limits for
zeroScaleNominal
0.0 VDC
0.1 VDC
SIGMAZ Noise Z Standard deviation of the
source closed signal
<0.00001 VDC
2
AIS AIS Nominal
AIS Range
Expected voltage output of the
electrometer with nothing in
the gap
Plus and minus limits for
fullScaleNominal
3.0-9.0 VDC
0.5 VDC
SIGMAB Noise F Standard deviation of the
electrometer output signal
with nothing in the gap.
<0.4 VDC
2
DIRT DIRT Nominal
DIRT Range
One half of the allowed dirt
accumulation permitted for
the sensor in gsm
Equal to the dirtNominal entry
TLK/STLK - 18.5 GSM ± 21.25
TLP/STLP - 4.5 GSM ± 5.5
TLS - 45 GSM ± 5 5